QE Coverage

QECoverage.png

Overview

QECoverage is a small planning tool to allow inelastic neutron scattering users to determine graphically the momentum / energy transfer limits for a given spectrometer and fixed energy.

Options

To use, select the geometry (direct or indirect) and instrument from the tabs and drop-down menu. For direct geometry spectrometers, an incident energy (Ei) in meV must be given; for indirect geometry machines, the final energies are fixed at a few values which must be selected from the drop-down menu. Only positive values of Ei are permitted. Several Q-E trajectories can be plotted simultaneously by providing several Ei values separated by commas.

To overplot the calculated Q-E trajectories, set the “Plot Over” check box.

The plot range is from Emin to Ei for direct, and from -Ef to Emax for indirect geometry spectrometers, and Emin and Emax may be inputted in the appropriate text box. If the input for Emin is invalid or bigger than min(Ei), then Emin is automatically set to Emin = -max(Ei)/2 for direct (indirect) geometry.

You can choose to create a 1D Mantid workspace for latter plotting using the “Create Workspace” check box.

In addition, for HYSPEC, where the detector bank can be rotated, an additional parameter S2 denoting the scattering angle of the center of the detector bank is required. The detectors extend over 60 degrees, so the two theta limits will be taken from max([0, abs(s2)-30]) to abs(s2)+30.

The Q(E) trajectory is calculated from the kinematic conditions:

\(\Delta E = E_i - E_f\)

and

\(\frac{\hbar^2 \mathbf{Q}^2}{ 2 m_n } = E_f + E_i - 2 \sqrt{E_i E_f} \cos{2\theta}\)

which is then solved for known \(E_i\) (\(E_f\)) for direct (indirect) geometry.